Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults

Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults

Authors

  • Jeanette Janaina Jaber Lucato Department of Physiotherapy, Centro Universitário São Camilo, Ipiranga, Sao Paulo
  • Thiago Marraccini Nogueira da Cunha Department of Physiotherapy, Centro Universitário São Camilo, Ipiranga, Sao Paulo
  • Sara Solange Oliveira Costa Rocha Department of Physiotherapy, Centro Universitário São Camilo, Ipiranga, Sao Paulo
  • Fernanda Maria Palmieri de Carvalho Department of Physiotherapy, Centro Universitário São Camilo, Ipiranga, Sao Paulo
  • Daniele Cristina Botega Department of Physiotherapy, Centro Universitário São Camilo, Ipiranga, Sao Paulo
  • Jamili Anbar Torquato Physiotherapist, São Paulo
  • Ana Cristina Gimenes Department of Physiotherapy, Centro Universitário São Camilo, Ipiranga, Sao Paulo
  • Renato Fraga Righetti Department of Rehabilitation, Hospital Sírio Libanês, São Paulo

Keywords:

Heat and moisture exchanger, Respirometer, Maximum inspiratory pressure, Maximum expiratory pressure, Manovacuometer

Abstract

Background: The use of evaluation tools such as the manovacuometer and respirometer is frequent and disinfection is usually limited to the external surfaces, which is insufficient and raises concerns because of the potential spread of infectious diseases. Hydrophobic heat and moisture exchangers (HME) are used in mechanical ventilation and have microbiological filters, which can possibly reduce contamination, increasing the safety of related procedures. It is unknown, however, if the addition of an exchanger affects the measurements obtained. Aim of this study was to verify if the use of an HME interferes in maximal inspiratory and expiratory pressures assessed using the manovacuometer and vital capacity evaluated using the respirometer in healthy adults.

Methods: A controlled transversal trial was carried out. Twenty healthy young adults were included in the study. Vital capacity by respirometer and, maximal inspiratory pressure (MIP) and maximal expiratory pressure (MEP) were assessed with and without the use of HME.

Results: No significant difference was found between the values pre and post HME use in vital capacity measurements: (3878.8 ± 202.2 mL vs. 3925.5 ± 206.0 mL, p = 0.116) and the respiratory muscle strength measurements: MIP (−99.0 ± 8.9 vs −95.5 ± 9.0 cm H2O, p = 0.149) and MEP (92.5 ± 7.5 vs 92.5 ± 7.7 cm H2O, p = 1.0) respectively.

Conclusion: We conclude that the use of HME does not modify the lung volumes or respiratory muscle strength, and can be used in order to reduce the occurrence of pulmonary infection.

Downloads

Published

19-12-2015

Issue

Section

Original Research Articles

How to Cite

1.
Jaber Lucato JJ, Marraccini Nogueira da Cunha T, Solange Oliveira Costa Rocha S, Palmieri de Carvalho FM, Botega DC, Anbar Torquato J, et al. Influence of heat and moisture exchanger use on measurements performed with manovacuometer and respirometer in healthy adults. Multidiscip Respir Med [Internet]. 2015 Dec. 19 [cited 2024 Jul. 4];11(1). Available from: https://mrmjournal.org/index.php/mrm/article/view/261